Paramagnetic oxygen analyzer (ZKG)

No need for reference gas
Ideal for measuring oxygen in flammable gas

  • Dumbbell type paramagnetic system requires no reference gas
  • Suited for oxygen in flammable gas
  • Interference effect is negligible
  • Japanese type approval

Specifications

Principle Paramagnetic (dumbbell type)
Range 0 to 10, 21, 25, 50, 100 vol% (as specified)
Measurement range 1 range or 2 ranges
Repeatability ≤ ±0.5% FS
Zero/span drift ≤ ±2% FS per week
Linearity ≤ ±1.0% FS
Response time ≤ 15 seconds (for 90% response)
Measured value output 4–20 mA DC, 0–1 V DC, or 0–10 mV DC
Contact input and output None
Display 4-digit LED
Gas inlet/outlet Rc1/4
Power supply voltage 85–264 V AC, 50/60 Hz
Dimensions 240 (H) × 192 (W) × 234 (D) mm
Mounting method Panel mount
Finish color Mansell 10Y7.5/0.5 semi gloss
Weight Approx. 5 kg

Paramagnetic oxygen analyzer (ZKG)